Bio-Rad Unity Alternatives for QC Peer Data

Vendor documents read

For the lab director or lab manager who sends control results to Bio-Rad's Unity Interlaboratory Program and wants to know what else is out there. Unity is the reference column; the four alternatives are the peer programs of three other control makers and of one analyzer maker. We read Bio-Rad's user guide and sample report and each vendor's brochures and product pages, beside 42 CFR 493. Daily QC software in general is compared in lab QC software.

The verdict by use case

Best for peer data with no monthly cut-off

Randox Acusera 24•7

Randox states peer data updated live, no submission deadlines and filters by method, instrument and reagent supplier.

Best for a lab network with its own peer group

Thermo Scientific LabLink

Thermo states custom peer groups for a trusted network of labs, beside the global peer data.

Best for a hematology bench on Sysmex

Sysmex BeyondCare Quality Monitor

The analyzer maker states real-time peer group evaluation and calibration verification each time QC runs.

Frosted-glass render of three control charts side by side under glass with control vials above
A1
Three control charts side by side: the view a peer program gives of one lab beside others on the same lot.

Peer program by feature

Each row is a question a lab asks before it moves its peer data: how the group is drawn, how results get in, how the report reads and how it relates to PT. CLIA requires none of these features, so only the last row carries a cite. Rows A and L repeat two rows of our QC software comparison, with the same statuses.

  • verified against a document
  • stated by the vendor
  • not stated

Scroll sideways for more vendors

QC peer-comparison programs, Bio-Rad Unity first
Requirement1Bio-Rad Unity2Thermo Scientific LabLink3Randox Acusera 24•74Technopath IAMQC Peer5Sysmex BeyondCare Quality Monitor
AShows peer-group statistics for the same control lotBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: stated by the vendor
BSplits the peer group by instrument modelBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: stated by the vendor
CReports a method group beside the peer groupBio-Rad Unity: verified against a documentThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated
DKeeps the reagent in the peer-group definitionBio-Rad Unity: verified against a documentThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated
EPrints how many labs sit in the peer groupBio-Rad Unity: verified against a documentThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: not statedSysmex BeyondCare Quality Monitor: not stated
FTakes results typed in on screenBio-Rad Unity: verified against a documentThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated
GImports an export file from the analyzer, LIS or middlewareBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: not statedSysmex BeyondCare Quality Monitor: not stated
HFeeds results automatically from the analyzer, LIS or middlewareBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated
IIssues a monthly peer reportBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated
JShows peer data on screen between monthly reportsBio-Rad Unity: stated by the vendorThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: stated by the vendor
KFlags the tests that need a supervisor's or director's reviewBio-Rad Unity: verified against a documentThermo Scientific LabLink: stated by the vendorRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: stated by the vendor
LAccepts control lots from other manufacturersBio-Rad Unity: verified against a documentThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: not statedSysmex BeyondCare Quality Monitor: not stated
MSold as a check between PT events, never as PT itself§ 493.801Bio-Rad Unity: stated by the vendorThermo Scientific LabLink: not statedRandox Acusera 24•7: stated by the vendorTechnopath IAMQC Peer: stated by the vendorSysmex BeyondCare Quality Monitor: not stated

Program by program

Bio-Rad UnityColumn 1

What it is

The Unity Interlaboratory Program: Bio-Rad's peer program, fed from UnityWeb or Unity Real Time. Each test is defined by analyte, instrument or kit, reagent, method, unit and temperature, and the monthly report sets your mean beside a peer group and a method group.

Who it fits

The reference point on this page. A lab already on Bio-Rad controls, because interlaboratory reports are available only for Bio-Rad control products, and not for a test set up under the code Other.

What a peer program should do, and whether it does

  • AShows peer-group statistics for the same control lot

    verified, source

  • BSplits the peer group by instrument model

    verified, source

  • CReports a method group beside the peer group

    verified, source

  • DKeeps the reagent in the peer-group definition

    verified, source

  • EPrints how many labs sit in the peer group

    verified, source

  • FTakes results typed in on screen

    verified, source

  • GImports an export file from the analyzer, LIS or middleware

    verified, source

  • HFeeds results automatically from the analyzer, LIS or middleware

    verified, source

  • IIssues a monthly peer report

    verified, source

  • JShows peer data on screen between monthly reports

    stated, vendor page; check with the vendor

  • KFlags the tests that need a supervisor's or director's review

    verified, source

  • LAccepts control lots from other manufacturers

    verified, source

  • MSold as a check between PT events, never as PT itself§ 493.801

    stated, vendor page; check with the vendor

Bio-Rad Unity website

What it is

Thermo Fisher's peer program for its MAS controls, sold as LabLink 360 and the older LabLink xL. Its LabLink xL brochure states real-time peer data, custom peer groups, file upload or a hands-free feed, and no submission cut-off.

Who it fits

A lab on MAS controls, or a group of affiliated labs that wants to compare against its own network as well as the global peer group. Ask which product, 360 or xL, your account will be on.

What a peer program should do, and whether it does

  • AShows peer-group statistics for the same control lot

    stated, vendor page; check with the vendor

  • BSplits the peer group by instrument model

    stated, vendor page; check with the vendor

  • CReports a method group beside the peer group

    not stated; check with the vendor

  • DKeeps the reagent in the peer-group definition

    not stated; check with the vendor

  • EPrints how many labs sit in the peer group

    not stated; check with the vendor

  • FTakes results typed in on screen

    not stated; check with the vendor

  • GImports an export file from the analyzer, LIS or middleware

    stated, vendor page; check with the vendor

  • HFeeds results automatically from the analyzer, LIS or middleware

    stated, vendor page; check with the vendor

  • IIssues a monthly peer report

    stated, vendor page; check with the vendor

  • JShows peer data on screen between monthly reports

    stated, vendor page; check with the vendor

  • KFlags the tests that need a supervisor's or director's review

    stated, vendor page; check with the vendor

  • LAccepts control lots from other manufacturers

    not stated; check with the vendor

  • MSold as a check between PT events, never as PT itself§ 493.801

    not stated; check with the vendor

Thermo Scientific LabLink website

Randox Acusera 24•7Column 3

What it is

Randox's cloud QC program with live peer statistics for its Acusera third-party controls. Peer data can be filtered by lot, analyte, method, instrument and reagent supplier, and results go in by hand, by file or through the optional Connect module.

Who it fits

A small lab that wants no monthly deadline and nothing installed. Randox says the software's internal functions work with any maker's QC material; ask whether peer data does too.

What a peer program should do, and whether it does

  • AShows peer-group statistics for the same control lot

    stated, vendor page; check with the vendor

  • BSplits the peer group by instrument model

    stated, vendor page; check with the vendor

  • CReports a method group beside the peer group

    stated, vendor page; check with the vendor

  • DKeeps the reagent in the peer-group definition

    stated, vendor page; check with the vendor

  • EPrints how many labs sit in the peer group

    stated, vendor page; check with the vendor

  • FTakes results typed in on screen

    stated, vendor page; check with the vendor

  • GImports an export file from the analyzer, LIS or middleware

    stated, vendor page; check with the vendor

  • HFeeds results automatically from the analyzer, LIS or middleware

    stated, vendor page; check with the vendor

  • IIssues a monthly peer report

    stated, vendor page; check with the vendor

  • JShows peer data on screen between monthly reports

    stated, vendor page; check with the vendor

  • KFlags the tests that need a supervisor's or director's review

    stated, vendor page; check with the vendor

  • LAccepts control lots from other manufacturers

    stated, vendor page; check with the vendor

  • MSold as a check between PT events, never as PT itself§ 493.801

    stated, vendor page; check with the vendor

Randox Acusera 24•7 website

Technopath IAMQC PeerColumn 4

What it is

Technopath's real-time peer program for its Multichem controls, part of the IAMQC family (its daily QC software is IAMQC Infinity, a separate product). Its peer group is instruments of the same type, on the same control lot and the same method; reports can be emailed monthly.

Who it fits

A lab on Multichem controls that wants a test-system peer and a method peer side by side, plus SDI and CVI exception reports, without running a server.

What a peer program should do, and whether it does

  • AShows peer-group statistics for the same control lot

    stated, vendor page; check with the vendor

  • BSplits the peer group by instrument model

    stated, vendor page; check with the vendor

  • CReports a method group beside the peer group

    stated, vendor page; check with the vendor

  • DKeeps the reagent in the peer-group definition

    stated, vendor page; check with the vendor

  • EPrints how many labs sit in the peer group

    not stated; check with the vendor

  • FTakes results typed in on screen

    stated, vendor page; check with the vendor

  • GImports an export file from the analyzer, LIS or middleware

    not stated; check with the vendor

  • HFeeds results automatically from the analyzer, LIS or middleware

    stated, vendor page; check with the vendor

  • IIssues a monthly peer report

    stated, vendor page; check with the vendor

  • JShows peer data on screen between monthly reports

    stated, vendor page; check with the vendor

  • KFlags the tests that need a supervisor's or director's review

    stated, vendor page; check with the vendor

  • LAccepts control lots from other manufacturers

    not stated; check with the vendor

  • MSold as a check between PT events, never as PT itself§ 493.801

    stated, vendor page; check with the vendor

Technopath IAMQC Peer website

Sysmex BeyondCare Quality MonitorColumn 5

What it is

The analyzer maker's own QC service for Sysmex XN-L, XN and XQ hematology analyzers. Sysmex states real-time peer group evaluation, and calibration verification each time QC is analyzed with Sysmex's own control material, such as XN CHECK.

Who it fits

A physician office lab or small lab whose hematology runs on a Sysmex analyzer. It covers that bench only, so chemistry still needs a program of its own.

What a peer program should do, and whether it does

  • AShows peer-group statistics for the same control lot

    stated, vendor page; check with the vendor

  • BSplits the peer group by instrument model

    stated, vendor page; check with the vendor

  • CReports a method group beside the peer group

    not stated; check with the vendor

  • DKeeps the reagent in the peer-group definition

    not stated; check with the vendor

  • EPrints how many labs sit in the peer group

    not stated; check with the vendor

  • FTakes results typed in on screen

    not stated; check with the vendor

  • GImports an export file from the analyzer, LIS or middleware

    not stated; check with the vendor

  • HFeeds results automatically from the analyzer, LIS or middleware

    not stated; check with the vendor

  • IIssues a monthly peer report

    not stated; check with the vendor

  • JShows peer data on screen between monthly reports

    stated, vendor page; check with the vendor

  • KFlags the tests that need a supervisor's or director's review

    stated, vendor page; check with the vendor

  • LAccepts control lots from other manufacturers

    not stated; check with the vendor

  • MSold as a check between PT events, never as PT itself§ 493.801

    not stated; check with the vendor

Sysmex BeyondCare Quality Monitor website

How each program builds a peer group

A peer report is only as useful as the match behind it. Unity defines every test by six parameters: analyte, instrument or kit, reagent type, method, unit of measure and temperature. With the lab and lot numbers, these “uniquely identify your QC data in the Unity Interlaboratory Program”, says Bio-Rad's Unity Real Time guide. Its monthly report then sets your mean beside two groups for each level: the peer group and a wider method group.

The alternatives draw the line in much the same place. Technopath's peer group holds “instruments of the same type from other laboratories using the same lot of QC material and using the same method” (IAMQC Peer brochure). Randox lets you filter its peer data by lot, analyte, method, instrument and reagent supplier. Thermo's LabLink xL adds custom peer groups, so a network of labs can compare against itself as well as the world.

Bigger is not always better. A Thermo SmartNote written by Sten Westgard runs the numbers: “An increase from 10 to 15 peers will get 14.8% closer to the mean, while an increase from 50 to 100 only gets 8.4% closer” (peer group comparisons). A few dozen labs on your exact analyzer, reagent and lot can tell you more than several hundred that share only a method.

Takeaway

Before you sign, ask each vendor how many labs report your analyzer, reagent and control lot today. The global participant count on the brochure is not the number that lands on your report.

Reading SDI and CVR on a peer report

Unity, Acusera 24•7 and IAMQC Peer each print two numbers per level (Randox and Technopath call the second one CVI). The standard deviation index (SDI) is your mean minus the group mean, divided by the group SD: how far off you sit. The coefficient of variation ratio (CVR) is your CV divided by the group CV: how noisy you are. Bio-Rad's QC statistics workbook calls an SDI of 1.49 or less acceptable, 1.5 to 1.99 a reason to investigate and 2.0 or more “generally considered to be unacceptable”; a CVR above 1.5 calls for investigation and above 2.0 for corrective action.

GGT level 2 from Bio-Rad's sample monthly report, with SDI and CVR worked out
Compared withMean (U/L)SDCV %LabsSDICVR
Your lab98.293.003.0
Peer group98.713.363.4234−0.120.9
Method group78.377.559.67212.640.3

The figures are GGT level 2 from Bio-Rad's sample monthly evaluation, a lab on a Siemens Dimension. Against its peer group the lab sits at an SDI of −0.12; against the method group, at 2.64, and the report prints a warning. Same lab, same month, same results. The method group pools labs on many analyzers that share the method, and its mean sits 20 U/L lower. Read the peer column first, and treat a method-only flag as a question about the method group before you treat it as a fault in your analyzer.

These limits are Bio-Rad's guidance, not CLIA's. CLIA sets no SDI limit; your written QC procedure decides what triggers an investigation.

Getting results in: by hand, by file or by feed

Peer data is only as current as your last submission, and there are three ways to make one. A lab with one analyzer can type results in: Unity records the initials of “the user who manually entered the data”, and Technopath accepts individual results or summary statistics (mean, SD and number of points) entered online. Randox's EDI route imports an export file from the analyzer, the laboratory information system (LIS) or middleware, after a one-time mapping. And the four control makers' programs each offer an automatic feed: UnityConnect, LabLink Auto-Connectivity, Acusera 24•7 Connect and IAMQC Transfer.

Unity's guide shows what the feed does on the bench. A hardware kit “taps the data exported from the laboratory instrument to the Laboratory Information Systems (LIS) without interrupting data flow”, and rows that cannot be matched land in a rejection log. The commonest cause is a code change: Bio-Rad reviews its analyte codelists and sends them out monthly, and a test whose setup no longer matches must be remapped before it imports again. A test parked under the code “Other” never reaches the interlaboratory report at all.

Takeaway

Whichever program you pick, ask who sees rejected imports and how often. A feed that silently drops one test leaves a gap in the peer report that nobody notices until month end.

Monthly reports, live peer data, and what the director signs

Unity's peer program runs on a monthly cycle: Bio-Rad says UnityWeb “enables participation in the monthly Unity Interlaboratory Program” and publishes a submission calendar. Technopath explains the difference plainly: a classic interlaboratory program builds its reports “at the end of the month once all data has been submitted”, while a real-time one uses everything submitted up to the moment you look (IAMQC Peer brochure). Thermo's LabLink xL and Randox's Acusera 24•7 both state that there is no submission cut-off at all.

Live data helps the bench when a control drifts on a Tuesday. The lab director cares about the monthly pack. CLIA asks the lab to “document all control procedures performed” (§ 493.1256(g)), and if your written QC procedure says the director reviews peer data each month, the signed report is the proof. Bio-Rad's sample monthly evaluation opens with the tests that “may require investigation or review”, which is the page a director actually reads. Ask every alternative for its equivalent, on your own data.

One more duty sits nearby. When the same test runs on two instruments, two methods or two sites, CLIA wants a system “that twice a year evaluates and defines the relationship between test results” (§ 493.1281(a)). Unity's Data Analysis Grid can “compare the QC results of like instruments running the same QC”. That is a useful start, but the comparison and its acceptance limits are still the lab's to write and sign. The lab QC software comparison covers the daily rule-checking side.

Peer data is not proficiency testing

Every vendor on this page sells peer comparison as a way to catch a problem early, and three of them say so in PT terms: Bio-Rad says Unity “can increase confidence in proficiency surveys”, Randox that Acusera 24•7 bridges “the gap between IQC and EQA”, and Technopath that IAMQC Peer builds confidence “prior to proficiency survey events”. Bio-Rad's own workbook calls an interlaboratory program “fully complementary of any proficiency testing program”. Complementary is the right word.

Deficiency risk

Peer reports are not PT

Proficiency testing runs under Subpart H with a CMS-approved program, which must offer “at least three testing events at approximately equal intervals per year”. A peer report shows how your controls compare. It is not graded, and it does not meet the Subpart H duty for a regulated analyte. The CMS-approved PT programs are listed by subspecialty.

Each lab “must enroll in a proficiency testing (PT) program that meets the criteria in subpart I” and is approved by HHS (§ 493.801). For a test that Subpart I does not list, CLIA asks the lab to verify accuracy “at least twice annually” (§ 493.1236(c)). The rule does not name peer data as a way to do that, so ask your accreditor or state agency before you lean on a peer report for it. The two PT programs most small labs choose between are compared in CAP vs API proficiency testing.

Leaving Unity: the controls go with it

Peer data follows the control, not the software. Bio-Rad's guide says interlaboratory reports “are not available for non-Bio-Rad control products”. LabLink is built for Thermo's MAS controls, IAMQC Peer for Technopath's Multichem and Acusera 24•7 for Randox's Acusera range, and none of the three states peer data for another maker's lots. In practice, a new peer program means a new control contract, and the price of the two belongs on one sheet.

New controls also mean new statistics. CLIA wants a mean and SD “for each batch and lot number” (§ 493.1256(d)(10)(i)), and for an unassayed control those figures “must be established over time by the laboratory through concurrent testing” beside material that already has them ((iii)). So run the new controls alongside the last Bio-Rad lot before it expires, not after, and chart the new lot in a Levey-Jennings chart maker while its own figures build up.

Then keep the history. QC records stay for “at least 2 years” (§ 493.1105(a)(3)), and Bio-Rad's sample report notes that its peer group statistics “may not be used without the expressed written consent of Bio-Rad”. Export your own lab's results and keep the signed monthly reports as they are, rather than copying peer figures into the new program.

Takeaway

Time the switch to a control lot change, and book the parallel run before the old lot's last month.
Note

How this comparison was made

We read Bio-Rad's Unity Real Time quick guide, sample monthly report, QC statistics workbook and Unity pages, Thermo Fisher's LabLink xL brochure, LabLink 360 page and flyers, Randox's Acusera 24•7 page and brochure, Technopath's IAMQC pages and Peer brochure, Sysmex's BeyondCare Quality Monitor page and brochure, and 42 CFR 493 on eCFR, last on .

A filled well means we verified the claim against a document we could read and link: the vendor's own documentation or a regulator's record. An open ring means the vendor says so in its own material, but we found no document to check it against. A blank well means we found nothing either way, which does not prove the product lacks it.

Verdicts are editorial: no vendor can buy a verdict position or displace another vendor. This page carries no prices and no ratings; ask each vendor for a quote and a demo on your own test menu.

Only Bio-Rad published user documentation and a full sample report we could read without an account, so its column holds the only filled wells; the other columns rest on brochures and product pages. Rows A and L carry the same statuses as on our QC software comparison. Sysmex is the one analyzer maker included, because its program covers the hematology bench of labs that run its analyzers. Proficiency-testing schemes were left out: they are a different duty, covered on their own page.

Questions labs ask

Is the Unity Interlaboratory Program the same thing as Unity Real Time?

No. UnityWeb and Unity Real Time are the QC software a lab works in each day; the Interlaboratory Program is the peer comparison that both feed. Bio-Rad describes UnityWeb as the entry level that “enables participation in the monthly Unity Interlaboratory Program” (UnityWeb page).

Can I use the peer mean as the target for my control?

Not on its own. CLIA lets a lab use the stated value of an assayed control only if it “is for the methodology and instrumentation employed by the laboratory and is verified by the laboratory” (§ 493.1256(d)(10)(ii)); for an unassayed control the lab builds its own figures over time ((iii)). A peer mean is a strong check on the figure you set, and a large SDI against it is a reason to look again.

Does CLIA require a peer-comparison program?

No. Nothing in 42 CFR 493 names one; the CLIA duties are daily QC, its records and proficiency testing for regulated analytes. Labs join for early warning, and labs working toward ISO 15189 meet a standard that, as Bio-Rad and Technopath both quote it, says the laboratory “shall participate in interlaboratory comparison” programs.

Can Unity give me peer reports on another maker's controls?

No. Unity Real Time tracks non-Bio-Rad lots inside your lab, but Bio-Rad's user guide says interlaboratory reports “are not available for non-Bio-Rad control products”. None of the alternatives here states peer data for another maker's lots either, so ask before you mix control brands.

What is Unity Next Peer QC?

A newer web tool Bio-Rad lists as “a simple peer comparison tool”, with peer reports on demand rather than at month end. We found it on Bio-Rad's regional site, which notes that some products have limited regional availability, so ask your Bio-Rad representative whether it is offered to your lab (Unity Next Peer QC page).