Best Digital Pathology Software for CLIA Labs

Vendor documents read

For pathology groups, reference labs and hospital labs choosing the image management system they will sign out on. We read FDA's clearance records for each product, each vendor's own pages and the CLIA rules in 42 CFR 493, and checked what each system does against what a CLIA lab has to show. Choosing the AP LIS itself is a separate comparison.

The verdict by use case

Best for mixed scanners

PathAI AISight Dx

Its cleared-product page names FDA clearance with Leica, Hamamatsu and Roche scanners, so one viewer can cover a fleet bought over several years.

Best for keeping your LIS in charge

Indica Labs HALO AP

Its product page says it assembles the case from the barcode and LIS data and sends the report back, working beside the LIS rather than replacing it.

Best for hospitals on one imaging platform

Sectra Digital Pathology

Pathology opens in the same IDS7 workspace as radiology, so a tumor board sees the slides and the scans in one place.

Frosted-glass render of stained slides in a rack, one laid on a gridded glass tile that repeats its tissue larger
A1
A rack of stained slides and the tiled image each one becomes: the scanner makes the picture, the software decides whose case it is and who reads it.

Requirement by requirement

  • verified against a document
  • stated by the vendor
  • not stated

Scroll sideways for more vendors

Digital pathology software by vendor, read 25 October 2026
Requirement1PathAI AISight Dx2Indica Labs HALO AP3Sectra Digital Pathology4Proscia Concentriq AP-Dx5Roche navify Digital Pathology6Philips IntelliSite Pathology Solution
ACleared by FDA for primary diagnosis21 CFR 864.3700PathAI AISight Dx: verified against a documentIndica Labs HALO AP: verified against a documentSectra Digital Pathology: verified against a documentProscia Concentriq AP-Dx: verified against a documentRoche navify Digital Pathology: verified against a documentPhilips IntelliSite Pathology Solution: verified against a document
BConnects to the LIS for case data§ 493.1283(a)⁠(1)PathAI AISight Dx: stated by the vendorIndica Labs HALO AP: verified against a documentSectra Digital Pathology: not statedProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: verified against a documentPhilips IntelliSite Pathology Solution: not stated
CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)PathAI AISight Dx: not statedIndica Labs HALO AP: not statedSectra Digital Pathology: not statedProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: verified against a documentPhilips IntelliSite Pathology Solution: not stated
DOpens slides from other makers' scannersPathAI AISight Dx: verified against a documentIndica Labs HALO AP: stated by the vendorSectra Digital Pathology: stated by the vendorProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: not statedPhilips IntelliSite Pathology Solution: not stated
EReview and consults from outside the labPathAI AISight Dx: verified against a documentIndica Labs HALO AP: verified against a documentSectra Digital Pathology: stated by the vendorProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: verified against a documentPhilips IntelliSite Pathology Solution: stated by the vendor
FRuns image analysis inside the viewerPathAI AISight Dx: stated by the vendorIndica Labs HALO AP: verified against a documentSectra Digital Pathology: stated by the vendorProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: stated by the vendorPhilips IntelliSite Pathology Solution: stated by the vendor
GMoves older cases to cheaper storage by policyPathAI AISight Dx: not statedIndica Labs HALO AP: not statedSectra Digital Pathology: not statedProscia Concentriq AP-Dx: stated by the vendorRoche navify Digital Pathology: not statedPhilips IntelliSite Pathology Solution: not stated

Vendor by vendor

PathAI AISight DxColumn 1

What it is

A cloud image management system for anatomic pathology: case lists a supervisor can assign, a slide viewer, threaded annotations and a consult workflow. PathAI's research edition runs PathAI and partner algorithms.

Who it fits

A lab or group with scanners from more than one maker that wants one viewer for sign-out, and that sends and receives consults.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    stated, vendor page; check with the vendor

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    not stated; check with the vendor

  • DOpens slides from other makers' scanners

    verified, source

  • EReview and consults from outside the lab

    verified, source

  • FRuns image analysis inside the viewer

    stated, vendor page; check with the vendor

  • GMoves older cases to cheaper storage by policy

    not stated; check with the vendor

PathAI AISight Dx website

Indica Labs HALO APColumn 2

What it is

A digital pathology platform that builds each case from the slide barcode and LIS data and sends the finished report back to the LIS. Indica sells it in the cloud or on premises.

Who it fits

A pathology group that is keeping its current AP LIS and wants the images to follow it rather than a second system of record.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    verified, source

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    not stated; check with the vendor

  • DOpens slides from other makers' scanners

    stated, vendor page; check with the vendor

  • EReview and consults from outside the lab

    verified, source

  • FRuns image analysis inside the viewer

    verified, source

  • GMoves older cases to cheaper storage by policy

    not stated; check with the vendor

Indica Labs HALO AP website

Sectra Digital PathologyColumn 3

What it is

The pathology module of Sectra's enterprise imaging platform: whole-slide images, gross photos and radiology studies in one diagnostic workspace, IDS7, with DICOM support.

Who it fits

A hospital or health system that already reads radiology on Sectra, or wants pathology and radiology side by side for tumor boards.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    not stated; check with the vendor

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    not stated; check with the vendor

  • DOpens slides from other makers' scanners

    stated, vendor page; check with the vendor

  • EReview and consults from outside the lab

    stated, vendor page; check with the vendor

  • FRuns image analysis inside the viewer

    stated, vendor page; check with the vendor

  • GMoves older cases to cheaper storage by policy

    not stated; check with the vendor

Sectra Digital Pathology website

Proscia Concentriq AP-DxColumn 4

What it is

The cleared clinical edition of Proscia's Concentriq platform. Proscia says the platform opens slides from 45+ scanners and syncs both ways with Epic Beaker and most LIS. Agilent announced an agreement to acquire Proscia on 13 October 2026.

Who it fits

A reference lab or multi-site group scanning at volume that wants scanner choice and storage rules it can set by policy.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    stated, vendor page; check with the vendor

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    stated, vendor page; check with the vendor

  • DOpens slides from other makers' scanners

    stated, vendor page; check with the vendor

  • EReview and consults from outside the lab

    stated, vendor page; check with the vendor

  • FRuns image analysis inside the viewer

    stated, vendor page; check with the vendor

  • GMoves older cases to cheaper storage by policy

    stated, vendor page; check with the vendor

Proscia Concentriq AP-Dx website

Roche navify Digital PathologyColumn 5

What it is

The cloud (SaaS) version of Roche's uPath software. Roche lists it under Roche Digital Pathology Dx, the cleared system of VENTANA DP 200 and DP 600 scanners, uPath software and a named display. The LIS link runs through Roche's Pathology Lab Hub.

Who it fits

A lab buying VENTANA scanners and stainers that wants one vendor for the scanner, the viewer and the service contract.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    verified, source

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    verified, source

  • DOpens slides from other makers' scanners

    not stated; check with the vendor

  • EReview and consults from outside the lab

    verified, source

  • FRuns image analysis inside the viewer

    stated, vendor page; check with the vendor

  • GMoves older cases to cheaper storage by policy

    not stated; check with the vendor

Roche navify Digital Pathology website

Philips IntelliSite Pathology SolutionColumn 6

What it is

Philips's scanners (SG60, SG300) with its image management system. SG scanners write Philips's own iSyntax format; Philips says its SGi scanners add DICOM output but are research use only in the United States.

Who it fits

A lab that wants a scanner and viewer sold and cleared together, from the maker whose system is the oldest whole slide imaging entry in FDA's records, granted in April 2017.

What CLIA asks, and whether it meets it

  • ACleared by FDA for primary diagnosis21 CFR 864.3700

    verified, source

  • BConnects to the LIS for case data§ 493.1283(a)⁠(1)

    not stated; check with the vendor

  • CKeeps a log of who did what on each case§ 493.1283(a)⁠(4)

    not stated; check with the vendor

  • DOpens slides from other makers' scanners

    not stated; check with the vendor

  • EReview and consults from outside the lab

    stated, vendor page; check with the vendor

  • FRuns image analysis inside the viewer

    stated, vendor page; check with the vendor

  • GMoves older cases to cheaper storage by policy

    not stated; check with the vendor

Philips IntelliSite Pathology Solution website

Digital pathology software, the scanner and the LIS

A digital pathology lab runs three systems. The scanner turns a glass slide into a whole-slide image. The image management system, which is what vendors mean by digital pathology software, stores that image, opens it for the pathologist and holds the case list, annotations and consults. The laboratory information system stays the record of the case: accessioning, blocks, slides, the signed report and the bill.

This page compares the middle system. What the AP LIS has to do once slides are scanned, such as tracking which slides are still glass, is covered in our pathology LIS comparison. Some LIS vendors now build a viewer in; the products here are bought on their own and connect to the LIS you have.

What FDA clearance covers, and what it does not

FDA files digital pathology products under one rule, 21 CFR 864.3700, a class II device, with two product codes. PSY is a whole slide imaging system: scanner, software and display cleared together. QKQ is the viewing and management software alone. FDA's database listed 14 QKQ clearances on 25 October 2026, from more makers than the six on this page.

The rule asks every submission to name “the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.” So a clearance is for a stack, not a brand: this viewer, with these scanners, on these displays. PathAI's page lists the scanners its clearance covers, and Roche's names its two VENTANA scanners, its uPath software and one display model.

The intended-use text Proscia and Roche publish also limits the tissue: slides from formalin-fixed, paraffin-embedded (FFPE) tissue, and not frozen sections, cytology or non-FFPE hematopathology. A lab that reads frozens or Paps on screen is outside those clearances, whatever the viewer can open.

FDA 510(k) and De Novo records for the six compared vendors, read 25 October 2026
VendorFirst recordLatest record
PathAIK212361Novo, 2022-08-11, QKQK243391AISight Dx, 2025-06-26, QKQ
Indica LabsK232833HALO AP Dx, 2024-05-07, PSYK252762HALO AP Dx, 2025-11-25, QKQ
SectraK193054Sectra Digital Pathology Module, 2020-03-31, QKQK232208Sectra Digital Pathology Module (Version 3.3), 2024-04-16, PSY
ProsciaK230839Concentriq Dx, 2024-02-08, PSYK261380Concentriq AP-Dx, 2026-08-05, QKQ
Roche (Ventana)K232879Roche Digital Pathology Dx (VENTANA DP 200), 2024-06-14, PSYK242783Roche Digital Pathology Dx, 2024-12-17, PSY
PhilipsDEN160056Philips IntelliSite Pathology Solution, 2017-04-12, PSYK243871Philips IntelliSite Pathology Solution 5.1, 2025-03-06, PSY

Look a vendor up yourself in FDA's 510(k) database by product code, then open the decision summary and read its indications for use. That list is the configuration to buy.

Validating digital sign-out before the first case

A cleared stack still has to be proven in your lab. CLIA asks a lab that introduces an unmodified FDA-cleared test system to show, before it reports patient results, that it performs as the maker says § 493.1253(b)⁠(1). A changed stack, such as another scanner, a different display or a tissue type outside the indications, puts the lab under § 493.1253(b)⁠(2): it sets its own performance specifications. Roche's labeling puts the same duty on the reader: “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images” (Roche).

The College of American Pathologists publishes the method. Its guideline on validating whole slide imaging, updated in 2021, keeps the 60-case validation set from the 2013 original and recommends that agreement between the glass diagnosis and the digital one be above 95%. CAP says studies going past 60 cases did not improve agreement. CAP frames validation around each intended clinical use, so a lab reading both H&E surgical cases and IHC on screen should cover both. CAP's guideline page.

Takeaway

Write the validation into the contract: the vendor installs the cleared configuration, you run the case set, and go-live waits for your sign-off. The same plan works for any software on the lab software shortlist.

Digital pathology image analysis and AI apps

Every platform here runs image analysis inside the viewer: scoring a stain, counting cells or flagging a slide to read first. Most of it comes from partners. Proscia says 150 apps work with Concentriq, from makers such as Ibex, Mindpeak, Visiopharm and Aiforia; PathAI lists DeepBio, DoMore Diagnostics, Paige and Visiopharm; Sectra says 50+ pathology AI vendors are integrated with its platform.

An app being open inside a cleared viewer does not make the app cleared. Proscia says so in its own footnote: interoperability “does not confer diagnostic intended use”, and each app's status is set by its own maker (Proscia).

Caution

Research use only is a label, not a draft

Several vendors sell a research edition next to the cleared one. Proscia says Concentriq AP and its AI apps are research use only and Concentriq AP-Dx is the clinical edition; Roche marks its uPath image analysis algorithms research use only; Philips says the same of PIPS 6 and Ibex AI in the United States; its latest cleared record is version 5.1. Ask which edition the demo runs, and for each algorithm you plan to report from, ask for its own clearance or treat it as a test your lab validates.

Scanning does not retire the glass

CLIA's retention rule is written for slides, not images. Histopathology slides are kept at least 10 years and cytology slides at least 5 years from the date of examination, and pathology blocks at least 2 years § 493.1105(a)⁠(7). A scanned slide still goes to the slide room. The full set of clocks is on our pathology LIS page.

That may change. CMS's proposed rule CMS-3478-P, published 8 October 2026, would add “digital images, digital data” to the items a lab must retain and add “and digital images” to the cytology and histopathology lines, so a lab would keep both the slide and its image. CMS gives the reason: either may be needed to trace an error behind a complaint or a quality problem. Comments are due 7 December 2026, and nothing changes before a final rule. Read the proposed rule.

Takeaway

Price storage as if the proposal becomes final: a histopathology image kept for the same 10 years as its slide. Ask each vendor what that costs per year at your volume, and how a case comes back from cold storage.

Remote sign-out under CLIA today

Remote review rests on a CMS memo, not on the regulation. Memo QSO-23-15-CLIA, issued 11 May 2023, lets pathologists and other qualified staff review digital data, results and images away from the lab under the primary lab's certificate. CMS's own summary of its conditions includes these:

  • The director of the primary lab is responsible for all testing under its certificate, remote work included.
  • The primary lab is certified in the specialties and subspecialties of the work done remotely.
  • Each remote reviewer is on the primary lab's Form CMS-209, the Laboratory Personnel Report.
  • Reports should show the remote site location; a code in place of a home address is allowed if it is available on request.
  • The primary lab keeps all documentation of the remote work and gives CMS a list of remote staff on request.

Cytology is the exception. Under the revised memo of 23 September 2025, anyone reviewing cytology slide images away from the certified lab has needed a separate CLIA certificate for that location since 23 March 2026 (the revised memo). The proposed rule would turn the rest into a defined “virtual access”, still excluding cytology.

Two of those conditions are software questions. The viewer should record who opened and signed each case, and from where, which supports the testing record CLIA asks for: the identity of the personnel who performed the test § 493.1283(a)⁠(4). And the report has to carry the remote site, so check where the LIS takes that field from. Confirm your setup with your CMS state survey agency or accreditor before the first remote case.

Note

Proposed, not yet law

CMS-3478-P (91 FR 64480) would define virtual access as review through a secure connection to the primary lab's software, LIS, instruments or EHR. Until a final rule appears, QSO-23-15-CLIA and its revision are what a surveyor reads.

Questions to put to every vendor in the demo

  1. Show the FDA record. Which 510(k) covers the configuration you are quoting, and are our scanners and display models in its indications for use?
  2. Scan, then open. Scan a slide with a barcode from our LIS and show it landing in the right case with no one typing a case number.
  3. Show the log. Pull the record of who opened, annotated and signed out one case, with the time and place of each step.
  4. Show the edition. For every algorithm in the demo, is it cleared, or research use only?
  5. Restore a case. Bring back a case from the oldest storage tier and time it.
  6. Leave in writing. How do we export every image and annotation in an open format if we change vendor? Images may have to outlive the contract by years.
Note

How this comparison was made

We read FDA's 510(k) and De Novo records under product codes QKQ and PSY, 21 CFR 864.3700, each vendor's own product and solutions pages, the CLIA rules in 42 CFR 493 on eCFR, CAP's whole slide imaging validation guideline page and the Federal Register text of CMS's October 2026 proposed rule, last on .

A filled well means we verified the claim against a document we could read and link: the vendor's own documentation or a regulator's record. An open ring means the vendor says so in its own material, but we found no document to check it against. A blank well means we found nothing either way, which does not prove the product lacks it.

Verdicts are editorial: no vendor can buy a verdict position or displace another vendor. This page carries no prices and no ratings; ask each vendor for a quote and a demo on your own test menu.

A filled well means FDA's record, or the page for the vendor's cleared product, names the function. A ring means a solutions page, a platform-wide page or the page for a research-use edition names it. FDA lists more cleared makers than these six; we compared the ones whose current US product pages we read for this page.

Questions labs ask

Is digital pathology software the same as a pathology LIS?

No. The LIS is the record of the case, from accessioning to the signed report and the bill. Digital pathology software stores and shows the slide images and runs the review around them. They talk to each other through an interface. To choose the LIS side, see the best pathology LIS for anatomic pathology labs.

Do we have to keep the glass slides after scanning?

Yes. CLIA's retention periods are for the slides: histopathology slides at least 10 years and cytology slides at least 5 years from the date of examination (§ 493.1105). CMS has proposed adding the digital images to the retained items as well; that is a proposal, not yet a rule.

Can our pathologists sign out from home?

For most work, yes, under CMS memo QSO-23-15-CLIA and its conditions: the reviewer is on your Form CMS-209, the report shows the remote site or a code for it, and your lab keeps the records. Cytology image review away from the certified lab has needed its own CLIA certificate since 23 March 2026.

If the software is FDA-cleared, do we still validate it?

Yes. CLIA asks every lab to verify an FDA-cleared system's performance before reporting patient results (§ 493.1253(b)(1)), and to establish its own specifications if it changes the cleared configuration. CAP's 2021 guideline update describes a 60-case validation set.

Is digital pathology image analysis software cleared for diagnosis?

Check each algorithm on its own. Each app's status is set by its own maker, and running inside a cleared viewer does not change it. Several algorithms named on these vendors' pages are labeled research use only in the United States.